# ICS Advisory (ICSMA-18-235-01)
## BD Alaris Plus
Original release date: August 23, 2018
[Print Document](javascript:window.print\(\);)
[Tweet](https://twitter.com/share?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2FICSMA-18-235-01)
[Like Me](https://www.facebook.com/sharer.php?u=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2FICSMA-18-235-01)
[Share](http://www.addthis.com/bookmark.php?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2FICSMA-18-235-01)
### Legal Notice
All information products included in [https://us-cert.gov/ics](/ics) are
provided "as is" for informational purposes only. The Department of Homeland
Security (DHS) does not provide any warranties of any kind regarding any
information contained within. DHS does not endorse any commercial product or
service, referenced in this product or otherwise. Further dissemination of
this product is governed by the Traffic Light Protocol (TLP) marking in the
header. For more information about TLP, see [https://www.us-
cert.gov/tlp/](/tlp/).
* * *
## 1\. EXECUTIVE SUMMARY
* **CVSS v3 9.4**
* **ATTENTION:** Exploitable remotely/low skill level to exploit
* **Vendor:** Becton, Dickinson and Company (BD)
* **Equipment:** Alaris GS, Alaris GH, Alaris CC, Alaris TIVA
* **Vulnerability:** Improper Authentication
## 2\. RISK EVALUATION
Successful exploitation of this vulnerability may allow a remote attacker to
gain unauthorized access to various Alaris Syringe pumps and impact the
intended operation of the pump when it is connected to a terminal server via
the serial port. BD has determined the affected products are not sold within
the United States.
## 3\. TECHNICAL DETAILS
### 3.1 AFFECTED PRODUCTS
The following versions of Alaris Plus, medical syringe pumps, Versions 2.3.6
and prior, are affected:
* Alaris GS,
* Alaris GH,
* Alaris CC, and
* Alaris TIVA.
### 3.2 VULNERABILITY OVERVIEW
#### 3.2.1 [IMPROPER AUTHENTICATION
CWE-287](https://cwe.mitre.org/data/definitions/287.html)
The software does not perform authentication for functionality that requires a
provable user identity.
[CVE-2018-14786](http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-14786)
has been assigned to this vulnerability. A CVSS v3 base score of 9.4 has been
calculated; the CVSS vector string is
([AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H](https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H)).
### 3.3 BACKGROUND
* **CRITICAL INFRASTRUCTURE SECTORS:** Healthcare and Public Health
* **COUNTRIES/AREAS DEPLOYED:** European Union
* **COMPANY HEADQUARTERS LOCATION:** United States
### 3.4 RESEARCHER
Elad Luz of CyberMDX discovered this vulnerability, and BD reported this
vulnerability to NCICC.
## 4\. MITIGATIONS
BD reports that this vulnerability cannot be performed if the device is
connected to an Alaris Gateway Workstation docking station. Also, an attacker
cannot switch the device on remotely, and no PHI or PII can be accessed by
exploiting this vulnerability.
BD recommends the following mitigations and compensating controls in order to
reduce risk associated with this vulnerability:
* This attack utilizes a known vulnerability in terminal servers. Users who utilize terminal servers should understand that terminal server use is not supported.
* Users should ensure they are operating these devices in a segmented network environment or as a stand-alone device.
* Users should utilize connections via the Alaris Gateway Workstation docking station, which would inactivate the remote control feature.
For more information on BD's product security and vulnerability management,
contact their product security office at:
<http://www.bd.com/productsecurity>
NCCIC recommends users take defensive measures to minimize the risk of
exploitation of this vulnerability. Specifically, users should:
* Minimize network exposure for all control system devices and/or systems, and ensure that they are [not accessible from the Internet](/ics/alerts/ICS-ALERT-10-301-01).
* Locate control system networks and remote devices behind firewalls, and isolate them from the business network.
* When remote access is required, use secure methods, such as Virtual Private Networks (VPNs), recognizing that VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize that VPN is only as secure as the connected devices.
NCCIC reminds organizations to perform proper impact analysis and risk
assessment prior to deploying defensive measures.
NCCIC also provides a section for [control systems security recommended
practices](/ics/content/recommended-practices) on the ICS-CERT web page.
Several recommended practices are available for reading and download,
including [Improving Industrial Control Systems Cybersecurity with Defense-in-
Depth Strategies](/sites/default/files/recommended_practices/NCCIC_ICS-
CERT_Defense_in_Depth_2016_S508C.pdf).
Additional mitigation guidance and recommended practices are publicly
available on the [ICS-CERT website](/ics/) in the Technical Information Paper,
[ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation
Strategies](/ics/tips/ICS-TIP-12-146-01B).
Organizations observing any suspected malicious activity should follow their
established internal procedures and report their findings to NCCIC for
tracking and correlation against other incidents.
No known public exploits specifically target this vulnerability.
暂无评论