# ICS Medical Advisory (ICSMA-19-178-01)
## Medtronic MiniMed 508 and Paradigm Series Insulin Pumps
Original release date: June 27, 2019
[Print Document](javascript:window.print\(\);)
[Tweet](https://twitter.com/share?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2Ficsma-19-178-01)
[Like Me](https://www.facebook.com/sharer.php?u=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2Ficsma-19-178-01)
[Share](http://www.addthis.com/bookmark.php?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2Ficsma-19-178-01)
### Legal Notice
All information products included in [https://us-cert.gov/ics](/ics) are
provided "as is" for informational purposes only. The Department of Homeland
Security (DHS) does not provide any warranties of any kind regarding any
information contained within. DHS does not endorse any commercial product or
service, referenced in this product or otherwise. Further dissemination of
this product is governed by the Traffic Light Protocol (TLP) marking in the
header. For more information about TLP, see [https://www.us-
cert.gov/tlp/](/tlp/).
* * *
## 1\. EXECUTIVE SUMMARY
* **CVSS v3 7.1**
* **Vendor:** Medtronic
* **Equipment:** MiniMed 508 and Paradigm Series Insulin Pumps
* **Vulnerability:** Improper Access Control
## 2\. RISK EVALUATION
Successful exploitation of this vulnerability may allow an attacker with
adjacent access to one of the affected products to intercept, modify, or
interfere with the wireless RF (radio frequency) communications to or from the
product. This may allow attackers to read sensitive data, change pump
settings, or control insulin delivery.
## 3\. TECHNICAL DETAILS
### 3.1 AFFECTED PRODUCTS
The following Medtronic MiniMed Insulin Pumps are affected:
* MiniMed 508 pump - All versions
* MiniMed Paradigm 511 pump - All versions
* MiniMed Paradigm 512/712 pumps - All versions
* MiniMed Paradigm 712E pump - All versions
* MiniMed Paradigm 515/715 pumps - All versions
* MiniMed Paradigm 522/722 pumps - All versions
* MiniMed Paradigm 522K/722K pumps - All versions
* MiniMed Paradigm 523/723 pumps - Software versions 2.4A or lower
* MiniMed Paradigm 523K/723K pumps - Software versions 2.4A or lower
* MiniMed Paradigm Veo 554/754 pumps - Software versions 2.6A or lower
* MiniMed Paradigm Veo 554CM and 754CM models only - Software versions 2.7A or lower
### 3.2 VULNERABILITY OVERVIEW
### 3.2.1 [IMPROPER ACCESS CONTROL
CWE-284](https://cwe.mitre.org/data/definitions/284.html)
The affected insulin pumps are designed to communicate using a wireless RF
with other devices, such as blood glucose meters, glucose sensor transmitters,
and CareLink USB devices. This wireless RF communication protocol does not
properly implement authentication or authorization. An attacker with adjacent
access to one of the affected insulin pump models can inject, replay, modify,
and/or intercept data. This vulnerability could also allow attackers to change
pump settings and control insulin delivery.
[CVE-2019-10964](http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2019-10964)
has been assigned to this vulnerability. A CVSS v3 base score of 7.1 has been
calculated; the CVSS vector string is
([AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H](https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H)).
### 3.3 BACKGROUND
* **CRITICAL INFRASTRUCTURE SECTORS:** Healthcare and Public Health
* **COUNTRIES/AREAS DEPLOYED:** Worldwide
* **COMPANY HEADQUARTERS LOCATION:** Ireland
### 3.4 RESEARCHER
Based on earlier work performed by external researchers including Nathanael
Paul, Jay Radcliffe, and Barnaby Jack, and from recent work performed by
external researchers Billy Rios, Jonathan Butts, and Jesse Young, Medtronic
performed additional variant analysis and reported this vulnerability to
NCCIC.
## 4\. MITIGATIONS
Medtronic recommends U.S. patients who are currently using the affected
products talk to their healthcare provider about changing to a newer model
insulin pump with increased cybersecurity protection. Patients outside the
U.S. will receive a notification letter with instructions based on the country
where they live.
Medtronic recommends all patients take the cybersecurity precautions indicated
below.
CYBERSECURITY PRECAUTIONS RECOMMENDED FOR ALL PATIENTS:
* Maintain tight physical control of the pump and devices connected to the pump
* Do not share pump serial number
* Be attentive to pump notifications, alarms, and alerts
* Immediately cancel any unintended boluses (a single dose of insulin administered all at once)
* Do not connect to any third-party devices or use any software not authorized by Medtronic
* Disconnect CareLink USB devices from computers when not being used to download data from the pump
* Monitor blood glucose levels closely and act as appropriate
* Get medical help immediately when experiencing symptoms of severe hypoglycemia or diabetic ketoacidosis, or suspect an insulin pump settings, or insulin delivery have changed unexpectedly
Medtronic has released additional patient-focused information, at the
following location:
<https://www.medtronic.com/security>
Additionally, Medtronic will be sending a letter to all patients who are
current known users of these pumps further detailing the risks and defensive
measures.
NCCIC recommends users take defensive measures to minimize the risk of
exploitation of this vulnerability. Specifically, users should:
* Restrict system access and access to devices able to connect to the affected device to authorized personnel only and follow a least privilege approach.
* Where additional information is needed, refer to existing cybersecurity in medical device guidance issued by the FDA at the following location:
<https://www.fda.gov/MedicalDevices/DigitalHealth/ucm373213.htm>
NCCIC reminds organizations to perform proper impact analysis and risk
assessment prior to deploying defensive measures.
NCCIC also provides a section for [control systems security recommended
practices](https://ics-cert.us-cert.gov/content/recommended-practices) on the
ICS-CERT web page. Several recommended practices are available for reading and
download, including [Improving Industrial Control Systems Cybersecurity with
Defense-in-Depth Strategies](https://www.us-
cert.gov/sites/default/files/recommended_practices/NCCIC_ICS-
CERT_Defense_in_Depth_2016_S508C.pdf).
Additional mitigation guidance and recommended practices are publicly
available on the [ICS-CERT website](https://ics-cert.us-cert.gov/) in the
Technical Information Paper, [ICS-TIP-12-146-01B--Targeted Cyber Intrusion
Detection and Mitigation Strategies](https://ics-cert.us-cert.gov/tips/ICS-
TIP-12-146-01B).
Organizations observing any suspected malicious activity should follow their
established internal procedures and report their findings to NCCIC for
tracking and correlation against other incidents.
No known public exploits specifically target this vulnerability. This
vulnerability is not exploitable remotely. High skill level is needed to
exploit.