# ICS Medical Advisory (ICSMA-18-123-01)
## Philips Brilliance Computed Tomography (CT) System (Update A)
Original release date: October 10, 2019
[Print Document](javascript:window.print\(\);)
[Tweet](https://twitter.com/share?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2FICSMA-18-123-01)
[Like Me](https://www.facebook.com/sharer.php?u=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2FICSMA-18-123-01)
[Share](http://www.addthis.com/bookmark.php?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2FICSMA-18-123-01)
### Legal Notice
All information products included in [https://us-cert.gov/ics](/ics) are
provided "as is" for informational purposes only. The Department of Homeland
Security (DHS) does not provide any warranties of any kind regarding any
information contained within. DHS does not endorse any commercial product or
service, referenced in this product or otherwise. Further dissemination of
this product is governed by the Traffic Light Protocol (TLP) marking in the
header. For more information about TLP, see [https://www.us-
cert.gov/tlp/](/tlp/).
* * *
## 1\. EXECUTIVE SUMMARY
* **CVSS v3 8.4**
* **ATTENTION:** Low skill level to exploit
* **Vendor:** Philips
**\--------- Begin Update A Part 1 of 3 ----------**
* **Equipment:** Brilliance CT Scanners and MX8000 Dual EXP
**\--------- End Update A Part 1 of 3 ----------**
* **Vulnerabilities:** Execution with Unnecessary Privileges, Exposure of Resource to Wrong Sphere, Use of Hard-coded Credentials
## 2\. UPDATE INFORMATION
This updated advisory is a follow-up to the original advisory titled
ICSMA-18-123-01 Philips' Brilliance Computed Tomography (CT) System that was
published May 3, 2018, on the ICS webpage on us-cert.gov.
## 3\. RISK EVALUATION
Successful exploitation of these vulnerabilities may allow an attacker to
attain elevated privileges and access unauthorized system resources, including
access to execute software or to view/update files including patient health
information (PHI), directories, or system configuration. This could impact
system confidentiality, system integrity, or system availability. Philips has
received no reports of exploitation or incidents from clinical associated with
these vulnerabilities.
## 4\. TECHNICAL DETAILS
### 4.1 AFFECTED PRODUCTS
* Philips reports the vulnerabilities affect the following Brilliance CT Scanners:
* Brilliance 64 Versions 2.6.2 and below
* Brilliance iCT Versions 4.1.6 and below
* Brilliance iCT SP Versions 3.2.4 and below
* Brilliance CT Big Bore Versions 2.3.5 and below
**\--------- Begin Update A Part 2 of 3 ----------**
* MX8000 Dual EXP Systems ([CVE-2018-8857](http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8857) only)
**\--------- End Update A Part 2 of 3 ----------**
### 4.2 VULNERABILITY OVERVIEW
### 4.2.1 [EXECUTION WITH UNNECESSARY PRIVILEGES
CWE-250](https://cwe.mitre.org/data/definitions/250.html)
Philips Brilliance CT devices operate user functions from within a contained
kiosk in a Microsoft Windows operating system. Windows boots by default with
elevated Windows privileges, enabling a kiosk application, user, or an
attacker to potentially attain unauthorized elevated privileges. Also,
attackers may gain access to unauthorized resources from the underlying
Windows operating system.
[CVE-2018-8853](http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8853)
has been assigned to this vulnerability. A CVSS v3 base score of 6.1 has been
calculated; the CVSS vector string is
([AV:L/AC:L/PR:L/UI:R/S:U/C:H/I:L/A:L](https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:L/UI:R/S:U/C:H/I:L/A:L)).
### 4.2.2 [EXPOSURE OF RESOURCE TO WRONG SPHERE
CWE-668](https://cwe.mitre.org/data/definitions/668.html)
Vulnerabilities within the Brilliance CT kiosk environment could enable a
limited-access kiosk user or an unauthorized attacker to break-out from the
containment of the kiosk environment, attain elevated privileges from the
underlying Windows OS, and access unauthorized resources from the operating
system.
[CVE-2018-8861](http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8861)
has been assigned to this vulnerability. A CVSS v3 base score of 6.1 has been
calculated; the CVSS vector string is
([AV:L/AC:L/PR:L/UI:R/S:U/C:H/I:L/A:L](https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:L/UI:R/S:U/C:H/I:L/A:L)).
### 4.2.3 [USE OF HARD-CODED CREDENTIALS
CWE-798](https://cwe.mitre.org/data/definitions/798.html)
The software contains fixed credentials, such as a password or cryptographic
key, which it uses for its own inbound authentication, outbound communication
to external components, or encryption of internal data. An attacker could
compromise these credentials and gain access to the system.
[CVE-2018-8857](http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8857)
has been assigned to this vulnerability. A CVSS v3 base score of 8.4 has been
calculated; the CVSS vector string is
([AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H](https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H)).
### 4.3 BACKGROUND
* **Critical Infrastructure Sectors:** Healthcare and public Health
* **Countries/Areas Deployed:** Worldwide
* **Company Headquarters Location:** U.S.
### 4.4 RESEARCHER
Philips reported these vulnerabilities to NCCIC.
## 5\. MITIGATIONS
Philips has identified the following guidance and controlling risk
mitigations:
* Users should operate all Philips deployed and supported Brilliance CT products within Philips authorized specifications, including Philips approved software, software configuration, system services, and security configuration such as firewall operations.
* Philips also recommends users implement a comprehensive, multi-layered strategy to protect their systems from internal and external security threats, including restricting physical access of the scanner to only authorized personnel, thus reducing the risk of physical access being compromised by an unauthorized user.
Philips has also remediated hard-coded credential vulnerabilities for
Brilliance iCT 4.x and above versions. The Philips iCT-iPatient (v4.x) family
Instructions for Use (IFU) refers to the ability to manage credentials and is
accessible to entitled users from the [Philips
InCenter](https://incenter.medical.philips.com).
**\--------- Begin Update A Part 3 of 3 --------**
As the MX8000 Dual EXP has been out of support since 2017, Philips recommends
a replacement based on user need. Please contact a local sales organization.
Philips will be further assessing options for remediation with future product
introductions and/or upgrades across the CT & AMI modalities to address
identified security vulnerabilities. Users with questions about their specific
Brilliance CT or MX8000 Dual EXP products should [contact a Philips service
support team](http://www.usa.philips.com/healthcare/solutions/customer-
service-solutions).
**\--------- End Update A Part 3 of 3 --------**
The Philips' advisory is available at the following URL:
<http://www.philips.com/productsecurity> (link is external)
CISA recommends users take defensive measures to minimize the risk of
exploitation of these vulnerabilities. Specifically, users should follow these
measures:
* Do not click web links or open unsolicited attachments in email messages.
* Refer to [Recognizing and Avoiding Email Scams](https://www.us-cert.gov/reading_room/emailscams_0905.pdf) for more information on avoiding email scams.
* Refer to [Avoiding Social Engineering and Phishing Attacks](https://www.us-cert.gov/cas/tips/ST04-014.html) for more information on social engineering attacks.
* Locate medical devices behind firewalls, and isolate them from the business network.
CISA reminds organizations to perform proper impact analysis and risk
assessment prior to deploying defensive measures.
CISA also provides a section for [control systems security recommended
practices](https://www.us-cert.gov/ics/recommended-practices) on the ICS
webpage on [us-cert.gov](https://www.us-cert.gov/ics). Several recommended
practices are available for reading and download, including [Improving
Industrial Control Systems Cybersecurity with Defense-in-Depth
Strategies](https://www.us-
cert.gov/sites/default/files/recommended_practices/NCCIC_ICS-
CERT_Defense_in_Depth_2016_S508C.pdf).
Additional mitigation guidance and recommended practices are publicly
available on the [ICS webpage on us-cert.gov](https://www.us-cert.gov/ics) in
the Technical Information Paper, [ICS-TIP-12-146-01B--Targeted Cyber Intrusion
Detection and Mitigation Strategies](https://www.us-cert.gov/ics/tips/ICS-
TIP-12-146-01B).
Organizations observing any suspected malicious activity should follow their
established internal procedures and report their findings to CISA for tracking
and correlation against other incidents.
No known public exploits specifically target these vulnerabilities. These
vulnerabilities are not exploitable remotely.
暂无评论