# ICS Medical Advisory (ICSMA-19-318-01)
## Philips IntelliBridge EC40/80 (Update A)
Original release date: December 12, 2019
[Print Document](javascript:window.print\(\);)
[Tweet](https://twitter.com/share?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2Ficsma-19-318-01)
[Like Me](https://www.facebook.com/sharer.php?u=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2Ficsma-19-318-01)
[Share](http://www.addthis.com/bookmark.php?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2Ficsma-19-318-01)
### Legal Notice
All information products included in [https://us-cert.gov/ics](/ics) are
provided "as is" for informational purposes only. The Department of Homeland
Security (DHS) does not provide any warranties of any kind regarding any
information contained within. DHS does not endorse any commercial product or
service, referenced in this product or otherwise. Further dissemination of
this product is governed by the Traffic Light Protocol (TLP) marking in the
header. For more information about TLP, see [https://www.us-
cert.gov/tlp/](/tlp/).
* * *
## 1\. EXECUTIVE SUMMARY
* **CVSS v3 6.3**
* **ATTENTION:** Low skill level to exploit
* **Vendor:** Philips
* **Equipment:** IntelliBridge EC40 and EC80
* **Vulnerability:** Inadequate Encryption Strength
## 2\. UPDATE INFORMATION
This updated advisory is a follow-up to the original advisory titled
ICSMA-19-318-01 Philips IntelliBridge EC40/80 that was published November 14,
2019, on the ICS webpage on us-cert.gov.
## 3\. RISK EVALUATION
Successful exploitation of this vulnerability may allow an attacker
unauthorized access to the IntelliBridge EC40/80 hub and may allow access to
execute software, modify system configuration, or view/update files, including
unidentifiable patient data.
## 4\. TECHNICAL DETAILS
### 4.1 AFFECTED PRODUCTS
The following versions of IntelliBridge are affected:
* IntelliBridge EC40 Hub, all versions
* IntelliBridge EC80 Hub, all versions
The IntelliBridge EC40/80 is intended to transfer medical device data from one
format to another according to preset specifications. It performs data
transfer without controlling or altering the function or parameters of any
connected medical devices. The EC40/80 is not intended for use in connection
with active patient monitoring. If data is not received from the hub a
notification is generated on the PIC iX Central Station and customers are
advised to refer to the source medical device.
### 4.2 VULNERABILITY OVERVIEW
#### 4.2.1 [INADEQUATE ENCRYPTION STRENGTH
CWE-326](https://cwe.mitre.org/data/definitions/326.html)
The SSH server running on the affected products is configured to allow weak
ciphers. This could enable an unauthorized attacker with access to the network
to capture and replay the session and gain unauthorized access to the EC40/80
hub.
[CVE-2019-18241](http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2019-18241)
has been assigned to this vulnerability. A CVSS v3 base score of 6.3 has been
calculated; the CVSS vector string is
([AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L](https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L)).
### 4.3 BACKGROUND
* **CRITICAL INFRASTRUCTURE SECTORS:** Healthcare and Public Health
* **COUNTRIES/AREAS DEPLOYED:** Worldwide
* **COMPANY HEADQUARTERS LOCATION:** Netherlands
### 4.4 RESEARCHER
The Medical Technology Solutions team of NewYork-Presbyterian Hospital
reported this vulnerability to Philips.
## 5\. MITIGATIONS
Philips plans a new release to remediate this vulnerability by the end of Q3
2020.
As an interim mitigation to this vulnerability, Philips recommends the
following:
* Users should operate all Philips deployed and supported products within Philips authorized specifications, including Philips approved software, software configuration, system services, and security configuration.
* The medical device network should be logically or physically isolated from the hospital network as specified in the Philips Patient Monitoring System Security for Clinical Networks guide, available on Philips [InCenter](https://incenter.medical.philips.com/). There is no clinical requirement for these devices to communicate outside of the Philips clinical network.
**\--------- Begin Update A Part 1 of 1 ---------**
* SSH is not meant to be used for clinical purposes or for supporting this product. Utilizing network security controls (e.g., ACL on switch access port, firewall rules), users can block network access to/from the SSH port of the Philips EC40/EC80 device without affecting EC40/80 functionality. NOTE: Philips' network infrastructure devices (router/switches and firewalls) use SSH (Port 22) for access and maintenance. If these devices are part of the network fabric, care must be taken to ensure Port 22 is only blocked for EC40/80 and not for other devices that need it.
**\--------- End Update A Part 1 of 1 ---------**
Users with questions about their specific EC40/80 product should contact a
Philips service support team or regional service support.
The Philips advisory is available at the following URL:
<http://www.philips.com/productsecurity>
Please see the Philips product security website for the latest security
information for Philips products:
[https://www.philips.com/productsecurity](http://www.philips.com/productsecurity)
CISA recommends users take defensive measures to minimize the risk of
exploitation of this vulnerability. Specifically, users should:
* Restrict system access to authorized personnel only and follow a least privilege approach.
* Apply defense-in-depth strategies.
* Disable unnecessary accounts and services.
* Where additional information is needed, refer to existing cybersecurity in medical device guidance issued by the FDA that can be found at the following location:
<https://www.fda.gov/MedicalDevices/DigitalHealth/ucm373213.htm>
CISA reminds organizations to perform proper impact analysis and risk
assessment prior to deploying defensive measures.
CISA also provides a section for [control systems security recommended
practices](https://www.us-cert.gov/ics/recommended-practices) on the ICS
webpage on [us-cert.gov](https://www.us-cert.gov/ics). Several recommended
practices are available for reading and download, including [Improving
Industrial Control Systems Cybersecurity with Defense-in-Depth
Strategies](https://www.us-
cert.gov/sites/default/files/recommended_practices/NCCIC_ICS-
CERT_Defense_in_Depth_2016_S508C.pdf).
Additional mitigation guidance and recommended practices are publicly
available on the [ICS webpage on us-cert.gov](https://www.us-cert.gov/ics) in
the Technical Information Paper, [ICS-TIP-12-146-01B--Targeted Cyber Intrusion
Detection and Mitigation Strategies](https://www.us-cert.gov/ics/tips/ICS-
TIP-12-146-01B).
Organizations observing any suspected malicious activity should follow their
established internal procedures and report their findings to CISA for tracking
and correlation against other incidents.
No known public exploits specifically target this vulnerability. This
vulnerability is exploitable from an adjacent network.