# ICS Medical Advisory (ICSMA-20-049-01)
## Spacelabs Xhibit Telemetry Receiver (XTR)
Original release date: February 18, 2020
[Print Document](javascript:window.print\(\);)
[Tweet](https://twitter.com/share?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2Ficsma-20-049-01)
[Like Me](https://www.facebook.com/sharer.php?u=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2Ficsma-20-049-01)
[Share](http://www.addthis.com/bookmark.php?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2Ficsma-20-049-01)
### Legal Notice
All information products included in [https://us-cert.gov/ics](/ics) are
provided "as is" for informational purposes only. The Department of Homeland
Security (DHS) does not provide any warranties of any kind regarding any
information contained within. DHS does not endorse any commercial product or
service, referenced in this product or otherwise. Further dissemination of
this product is governed by the Traffic Light Protocol (TLP) marking in the
header. For more information about TLP, see [https://www.us-
cert.gov/tlp/](/tlp/).
* * *
## 1\. EXECUTIVE SUMMARY
* **CVSS v3 9.8**
* ATTENTION: Exploitable remotely/Low skill level to exploit/Public exploits known
* Vendor: Spacelabs
* Equipment: Xhibit Telemetry Receiver
* Vulnerability: Improper Input Validation
## 2\. RISK EVALUATION
A remote code execution vulnerability called BlueKeep
([CVE-2019-0708](https://cve.mitre.org/cgi-
bin/cvename.cgi?name=CVE-2019-0708)) exists within the Remote Desktop Protocol
(RDP) used by the Microsoft Windows operating systems listed below. An
attacker can exploit this vulnerability to perform remote code execution on an
unprotected system.
According to Microsoft, an attacker can send specially crafted packets to
operating systems with RDP enabled. After successfully sending the packets
the attacker could perform a number of actions, including adding accounts with
full user rights; viewing, changing, or deleting data; or installing programs.
This exploit, which requires no user interaction, must occur for successful
authentication.
BlueKeep is considered "worm-able" because malware exploiting this
vulnerability on a system could propagate to other vulnerable systems; thus a
BlueKeep exploit would be capable of rapidly spreading like the WannaCry
malware attacks of 2017.
## 3\. TECHNICAL DETAILS
### 3.1 AFFECTED PRODUCTS
The following versions and operating systems of Spacelabs Xhibit Telemetry
Receiver are affected:
* Xhibit Telemetry Receiver (XTR), Model number 96280, v1.0.2
* Arkon (99999), all versions - previously sold by Spacelabs, no longer a supported product
The following Microsoft Windows operating systems, including both 32- and
64-bit versions, as well as all Service Pack versions, are affected:
* Windows 2000
* Windows Vista
* Windows XP
* Windows 7
* Windows Server 2003
* Windows Server 2003 R2
* Windows Server 2008
* Windows Server 2008 R2
### 3.2 VULNERABILITY OVERVIEW
#### 3.2.1 [ IMPROPER INPUT VALIDATION
CWE-20](https://cwe.mitre.org/data/definitions/20.html)
The affected product is vulnerable to a remote code execution vulnerability
that exists in Remote Desktop Services (formerly known as Terminal Services)
when an unauthenticated attacker connects to the target system using RDP and
sends specially crafted requests. This vulnerability is pre-authentication and
requires no user interaction. An attacker who successfully exploited this
vulnerability could execute arbitrary code on the target system. An attacker
could then install programs; view, change, or delete data; or create new
accounts with full user rights. To exploit this vulnerability, an attacker
would need to send a specially crafted request to the target system's Remote
Desktop Service via RDP.
[CVE-2019-0708](http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2019-0708)
has been assigned to this vulnerability. A CVSS v3 base score of 9.8 has been
calculated; the CVSS vector string is
([AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H](https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H)).
### 3.3 BACKGROUND
* **CRITICAL INFRASTRUCTURE SECTORS:** Healthcare and Public Health
* **COUNTRIES/AREAS DEPLOYED:** Worldwide
* **COMPANY HEADQUARTERS LOCATION:** United States
### 3.4 RESEARCHER
The vulnerability was originally discovered by Microsoft. Spacelabs then
reported the vulnerability to CISA.
## 4\. MITIGATIONS
Spacelabs has determined the recommended remediation is to update to the
newest release v1.2.1 or later. All deployed XTR hardware appliances are
capable of update and should be updated.
Many Spacelabs products are appliances and users are not intended to perform
updates on them. Products or systems that are obsolete or are not able to be
patched may use this alternate mitigation step to help protect against
BlueKeep:
* Block Transmission Control Protocol (TCP) Port 3389 at the enterprise perimeter firewall. Because Port 3389 is used to initiate an RDP session, blocking it prevents an attacker from exploiting BlueKeep from outside the user's network. However, this will block legitimate RDP sessions and may not prevent unauthenticated sessions from being initiated inside a network.
Spacelabs also encourages users and administrators to review the [Microsoft
Security Advisory](https://portal.msrc.microsoft.com/en-US/security-
guidance/advisory/CVE-2019-0708) and the [Microsoft Customer Guidance for
CVE-2019-0708](https://support.microsoft.com/en-us/help/4500705/customer-
guidance-for-cve-2019-0708) and apply the appropriate mitigation measures as
soon as possible.
If you own an XTR device or have any questions about this security advisory,
please contact Spacelabs at 1-800-522-7025 and select 2 for technical support.
XTR is an appliance that has no user interface, so your service representative
can help you to determine the installed version of software on your XTR
product and will work to coordinate updates as needed.
For additional information about this vulnerability, please see the [Spacelabs
Security
Advisory](https://www.spacelabshealthcare.com/products/security/bluekeep-
cybersecurity-information-cve-2019-0708/).
CISA recommends users take defensive measures to minimize the risk of
exploitation of this vulnerability. Specifically, users should:
* Minimize network exposure for all medical devices and/or systems.
* Locate medical devices behind firewalls and isolate them where possible.
* Restrict system access to authorized personnel only and follow a least privilege approach.
* Apply defense-in-depth strategies.
* Disable any unnecessary accounts, protocols and services.
* Where additional information is needed, refer to [existing cybersecurity in medical device guidance](https://www.fda.gov/MedicalDevices/DigitalHealth/ucm373213.htm) issued by the FDA.
CISA reminds organizations to perform proper impact analysis and risk
assessment prior to deploying defensive measures.
CISA also provides a section for [control systems security recommended
practices](https://www.us-cert.gov/ics/recommended-practices) on the ICS
webpage on [us-cert.gov](https://www.us-cert.gov/ics). Several recommended
practices are available for reading and download, including [Improving
Industrial Control Systems Cybersecurity with Defense-in-Depth
Strategies](https://www.us-
cert.gov/sites/default/files/recommended_practices/NCCIC_ICS-
CERT_Defense_in_Depth_2016_S508C.pdf).
Additional mitigation guidance and recommended practices are publicly
available on the [ICS webpage on us-cert.gov](https://www.us-cert.gov/ics) in
the Technical Information Paper, [ICS-TIP-12-146-01B--Targeted Cyber Intrusion
Detection and Mitigation Strategies](https://www.us-cert.gov/ics/tips/ICS-
TIP-12-146-01B).
Organizations observing any suspected malicious activity should follow their
established internal procedures and report their findings to CISA for tracking
and correlation against other incidents.